| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Radiat Prot > Volume 23(1); 1998 > Article
Journal of Radiation Protection 1998;23(1):49-0.
중성자 개인선량계 상호비교
김봉환;김장렬;장시영;
Intercomparison Study of the Neutron Personnel Dosemeters
Kim, Bong-Hwan;Kim, Jang-Lyul;Chang, Si-Young;
ABSTRACT
Domestic intercomparison study of the neutron personnel dosemeters was performed for the first time in Korea. Thirteen types of neutron dosemeters from twelve institutions took part in this intercomparison study and the $D_2O$ moderated Cf-252 source of KAERI was used for irradiation. Eight of the fifteen dosemeters submitted by each participant were divided into two groups and each group was irradiated with different doses of the simulated mixed fields of neutron and gamma. The participants assessed their dosemeter reading in terms of the personal dose equivalent, Hp(10), for both neutron and gamma dose. The ratio of the reported dose equivalent to the delivered dose equivalent for comparison between participants ranged from 0.55 to 1.34 for neutron, from 0.54 to 1.32 for gamma and from 0.75 to 1.20 for total dose. This intercomparison results show that all dosemeter processors, especially for neutron category, are able to pass the personnel dosemeter performance test which shall be enforced according to the ordinance of the MOST, No. 96-6.
Editorial Office
#319, Hanyang Institute of Technology Bldg., 222 Wangsimni-ro, Seongdong-gu,Seoul, Republic of Korea
Tel: +82-2-2297-9775   Fax: +82-2-2297-9776
Email: managing.editor@jrpr.org
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © by Korean Association for Radiation Protection. Developed in M2PI