Journal of Radiation Protection 2001;26(2):73-0.
3" X 3" NaI 스펙트럼의 조사선량 변환계수 결정
이모성;
Determination of the exposure conversion coefficient for 3" X 3" NaI spectrum
Lee, M.S.;
ABSTRACT
In order to find the exposure conversion coefficients for 3"X3" NaI spectrum, we measured the exposure rates with the pressurized ion chamber at 29 different areas in the range of $4{sim}23{mu}R;h^{-1}$, and also measured the gamma spectra with 3"X3" and 4"X4" NaI detectors, simultaneously. The exposure conversion coefficient of the total energy method was determined using the linear relation between the measured exposure rate and the gamma spectrum energy. In order to find the exposure conversion coefficients of the energy band method, we applied the exposure conversion coefficients recommended by NCRP to the 4"X4" NaI spectra, and calculated the exposure rates due to $^{40}K,;^{238}U$, and $^{232}Th$ series respectively. Using the linearly proportional relation between the obtained $^{232}Th$ series exposure rate and peak area of 2614 keV that represents the $^{232}Th$ series, we obtained the exposure conversion coefficients for $^{232}Th$ series. We also determined the conversion coefficients for $^{238}U$ series and $^{40}K$ using a similar method.
Editorial Office
#319, Hanyang Institute of Technology Bldg., 222 Wangsimni-ro, Seongdong-gu,Seoul, Republic of Korea
Tel: +82-2-2297-9775   Fax: +82-2-2297-9776
Email: managing.editor@jrpr.org
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © by Korean Association for Radiation Protection. Developed in M2PI