A Measurement of Kerma and Absorbed Dose in Photon Fields
Kim, Sung-Hee;Shin, Seung-Aea;Chu, Sung-Sil;
ABSTRACT
Determination of the relation between the kerma(Kinetic Energy Released in Material) and the absorbed dose is one of the basic problems of dosimetry. Kerma and absorbed dose were measured for 6 MV X-ray from the high energy medical linear accelerator and $^{60}Co$ gamma-ray. The experimental results show that the absorbed dose in the transient equilibrium region practically coincide with the kerma in water and Al for $^{60}Co$. The maximum dose depths were $1.45g/cm^2$ for 6MV X-ray and $0.48g/cm^2;for;^{60}Co$ gamma-ray. The ratios of the absorbed dose at maximum build-up to the collision kerma at the surface, ($K^{att}$), were 0.949 for 6MV X-ray and 0.992 for $^{60}Co$ gamma-ray. No difference was found between water and Al when the standard field size was used. This results show that the dependence of $K^{att}$ on the material is very small.